Instruction manual
31 
Wait at least 3 minutes after 
turning the instrument on to allow sensor 
readings to stabilize fully before initiating any 
calibration procedures. Failure to wait three 
minutes before initiating calibration procedures 
may lead to inaccurate and potentially 
dangerous readings. If a sensor has just been 
replaced, the stabilization period may be longer 
(see the chart in section 5.1). 
(3) Slide the belt clip towards the rear of the 
instrument exposing the four buttons on the 
instrument keypad. 
(4) Verify that the instrument is in the Basic or 
Technician operating mode. To change 
operating modes simultaneously hold down the 
"+" and"-" keys. 
  Each time that the operating mode is changed, 
the LCD screen will briefly indicate the current 
operating mode. 
(5) Make sure the instrument is located in an area 
where the air is known to be fresh. 
(6) Press the keypad button marked "Cal". The 
fresh air calibration message will appear briefly 
on the instrument LCD. 
(7) The fresh air calibration message will be 
followed by the zero-adjustment screen. 
(8) Press "Cal" to automatically zero the instrument. 
An information screen will be briefly displayed 
while the adjustments are being made, and 
another when the adjustments have been 
completed. 
(9) Pressing the mode button causes the calibration 
values in the memory to remain unchanged 
from the last time a fresh air adjustment was 
made. An information screen will be displayed 
briefly which verifies that the zero values have 
not been changed. 
(10)  After completion of the zero adjustment the 
PhD Ultra will automatically return to the gas 
reading screen display. 
3.5.2  Span calibration using keypad 
buttons 
Span calibration procedures using buttons on the 
instrument keypad are most useful when calibration 
of only a single sensor is desired. Span calibration 
using the keypad buttons may only be done while in 
the Basic or Technician operating mode. 
(1) Turn the instrument on. 
(2) Slide the belt clip towards the rear of the 
instrument exposing the four buttons on the 
instrument keypad. 
(3) Verify that the instrument is in the Basic or 
Technician operating mode. To change 
operating modes simultaneously hold down the 
"+" and"-" keys. 
  Each time that the operating mode is changed, 
the LCD screen will briefly indicate the current 
operating mode. 
(4) Turn the instrument off by pressing the mode 
button for three full seconds until the "release 
button" message appears in the screen. 
(5) With the unit turned off, press and hold down 
the "Cal" button. 
(6) While holding down the "Cal" button, press the 
mode button to turn the PhD Ultra back on. The 
“Span Calibration” message will appear briefly 
on the screen. This message is followed by a 
screen showing the message “Cal = Span”. 
(7) Press the keypad button marked “Cal” to make 
a span adjustment. A screen will briefly display 
the message “Span Calibration; Mode = 
Cancel”. 
  Press the mode button at any time to cancel the 
calibration mode. 
  The “Span Calibration” screen will be followed 
by another showing the first sensor to be 
adjusted. 
(8) Attach the cylinder of calibration gas, regulator, 
short section of tubing and calibration adapter to 
the PhD Ultra as shown in Figure 3.4.2. 
Make sure the regulator, 
cylinder seating surfaces and threads are clean 
and dry before attaching the regulator to the 
cylinder of gas. Introduction of contaminants 
through the regulator fittings may alter or 
degrade the concentration of the gas contained 
in the cylinder and may lead to inaccurate and 
potentially dangerous gas readings. 










