Instruction Manual

Copyright © 2001 BioMedical Life Systems, Inc.
All Rights Reserved
Instructions
BIOSTIM® M7 3/18/22 REV. L
Transcutaneous
Electrical Nerve Stimulator
BioStim
®
M7
BioMedical Life Systems, Inc.
BioMedical Life Systems, Inc.
1954 Kellogg Avenue
Carlsbad, California 92008 USA
Tel: (800) 726-8367
Manufactured for and Distributed by:
www.bmls.com
LIMITED TWO-YEAR WARRANTY
BioMedical Life Systems, Inc. promises to the original
consumer-purchaser to repair or replace, at the option of
BioMedical Life Systems, Inc., any neurostimulator which
malfunctions or proves defective in materials or workmanship
under normal use for a warranty period from the date of
purchase. During the warranty period, BioMedical Life Systems,
Inc. will provide all labor and parts necessary to correct such
defects or malfunctions free of charge. It is the duty of the
consumer-purchaser to deliver the unit to a service facility of
the factory with a receipt of purchase. For further instructions,
please call 800-726-8367.
EXCLUSIONS:
This warranty shall not apply to damage resulting from fail-
ure to follow the operating instructions in the user manual,
accident, abuse, alteration, or disassembly by unauthorized
personnel. This warranty does not extend to accessory items
such as rechargeable batteries, electrodes, electrode and
leads wires which are not an integral part of the stimulator.
These items can be provided by your service representative,
but costs for repair or replacement will be the responsibility
of the consumer-purchaser. BioMedical Life Systems, Inc.
shall not be liable for incidental or consequential damages
resulting from the sale or use of the unit. Some states do
not allow the exclusion or limitation of incidental or conse-
quential damages, so the above limitation or exclusion may
not apply to you.
WARNING:
Only use accessories, electrodes, lead wires and batteries
approved by BioMedical Life Systems, Inc.
NO OTHER WARRANTIES:
This limited warranty is the only express warranty given by
BioMedical Life Systems, Inc. Implied warranties, including,
but not limited to, warranties of merchantability and tness
for a particular purpose are limited to the warranty period
set forth above. Some states do not allow limitations on how
long an implied warranty lasts, so the above limitation may
not apply to you. This warranty gives you specic legal rights,
and you may also have rights which vary from state to state.
Warranty
Device Maintenance
7
5 6
• Clean skin prior to placement of electrodes.
• Encircle the area of pain. The pads should never touch and
should be at least, 2" apart.
• Do not to place directly over a joint such as the knee, elbow
or ankle as movement can alter pad placement.
• When the pain is in a small area (e.g. calf pain): Place pads
in parallel on each side of the pain.
• When the pain overlaps a joint (e.g. elbow pain): Place
each pad on the muscle or soft tissue just above and below
the joint in a horizontal and parallel direction.
• When the pain is wide (e.g. between your shoulders below
the neck): Place the pads to the left and right side of your
spine in a vertical direction.
Place electrodes rmly to your body as you were shown by
your healthcare professional.
Warning: Only electrodes and lead wires authorized by the
device manufacturer should be used.
Electrode Placement Guidelines
Amplitude (4, 5)
Set the stimulation intensity by pressing the CH1 (4) and/
or CH2 Intensity (5) buttons to increase (+) or decrease (-)
to your own comfort level. Your selection will be displayed
on either side of the screen. There are 20 levels of intensity
available.
Timer (10)
Select the desired treatment time by pressing the timer but-
ton to continuous, 15, 30, or 60 minutes.
Battery (11)
In order to maintain the functional operation of the BioStim
®
M7 the batteries will have to be changed periodically. The
device is supplied with 2 AA Alkaline batteries.
To change batteries:
Before opening the battery compartment, check to make
sure that the device is switched o (6).
Open battery compartment (11).
Gently insert batteries by matching the +/– end of each
battery to the symbol found in the battery compartment.
Step 1
A. Insert battery and close battery cover (11).
B. Set mode (6) as recommended by your health professional.
C. Adjust settings (P.R.(8) & P.W.(7)) as recommended by
your health professional.
NOTE: Should a low frequency P.R. (5-20 Hz) TENS treatment
not yield the desired result a high frequency TENS treatment
Treatment Guidelines
PATIENT LOCK
The BioStim® M7 has a patient lock feature. When “turned
on”, this feature only allows the patient to adjust the intensity
controls and timer. No other parameters (mode, pulse rate,
pulse width) can be changed by the patient.
TURN ON
After the desired time has been selected, press the timer
button (10) and the mode button (6) simultaneously. Hold
both buttons until you see the clock diagram ashes on
screen. The patient lock feature is now activated.
TURN OFF
Press the timer button (10) and the Mode Button (6) simulta-
neously. Hold both buttons until the clock diagram stops
ashing. The patient lock feature is now deactivated.
PATIENT COMPLIANCE COUNTER
When turning o the patient lock feature, the amount of time
the patient has used the device is displayed in the number
of hours the device has been used will be displayed. Once
99 hours has been reached, a bar will appear next to the “1”
on the left amplitude display. Each bar indicates 100 hours
of use. A maximum of 1099 hours will be displayed.
CLEAR STORED HOURS
To clear the amount of usage hours stored, press and hold
the CH1 "DOWN arrow" and CH2 "DOWN arrow" simultane-
ously. Continue to hold until the patient compliance meter
resets to display 0 hours of use.
Patient Compliance Meter
should be applied in the range of 100-150 Hz. If no P.W.
recommendations are given, a pulse width setting of 50-100
μs are generally used.
D. Set the timer switch (10) to the desired treatment time.
Step 2
Attach lead wires to electrodes, and place electrodes rmly
to your body as you were shown by your health professional.
Step 3
Insert lead wires into device outlet (1,2).
Step 4
Slowly increase intensity by pressing the "UP arrow" sign
or decrease by pressing "DOWN arrow" on CH1 (4) and/or
CH2 (5) amplitude buttons. Increase amplitude as high as
possible without causing discomfort.
NOTE: The correct level of stimulation should feel comfort-
able to the patient and should never be set at levels that
cause discomfort.
Step 5 (End of treatment)
A. Turn o device.
B. Unplug lead wires.
C. Disconnect electrodes from lead wires.
D. Store electrodes according to the instructions on the
electrode package.
• Isopropyl alcohol is suitable for cleaning this device.
• Do not smoke or work with open ames when working
with ammable liquids!
• Stains and spots can be removed with a cleaning agent.
• Do not submerge this device in any liquid or use exces-
sive cleaning liquid when cleaning the surface area.
• Remove the batteries for long term storage.
• Damaged or leaking batteries should be disposed of in
the proper manner in accordance with local regulations.
• Use only BioMedical Life Systems, Inc. approved acces-
sories and electrodes. Non-approved accessories and elec-
trodes may damage your device and aect your warranty.