FIGURE 4-3. StimRouter remote control digital display and LEDs.
4-18 StimRouter User Manual
Radio Frequency Registration Error
“E” appears in the digital display.
Radio Frequency Registration Complete
“C” appears in the digital display.
Radio Frequency Registration in Progress
A green arch alternates in the digital display.
LEDs
User Patch Icon
Flashes red: faulty electrode contact or temperature error in the
external pulse transmitter
Glows red: component malfunction
Flashes yellow: low battery
Radio Frequency (RF) Icon
Flashes red: radio frequency error
Glows green for a few seconds: RF registration is complete
User Patch Icon
Digital Display
Remote Control Icon
Radio Frequency