User Manual

Table Of Contents
Operating Instructions 8-31
1 Locate the area where the pick-up electrode is implanted. If
necessary, palpate the area until the pick-up electrode can be felt.
2 Clean the skin above the pick-up electrode with an alcohol swab or
wet washcloth and then dry the skin area. If the area has been treated
with lotions or oils, then clean the skin with soap and water, rinse
well and dry. If necessary, remove excess body hair from the skin
area with scissors. Do not use a razor. A razor can irritate the skin.
3 Hold the user patch between the index finger and thumb, with the
gel electrodes facing toward the site of adhesion.
CAUTION:
Do not pinch or stretch the skin while adhering the user patch.
4 Orient the user patch so that the center of one of the gel electrodes is
positioned directly over the pick-up end of the lead. If the user patch
is not placed directly over the pick-up end of the lead, stimulation
may be uncomfortable or ineffective.
CAUTION:
Make certain that the adhesion site is free from any obstructions, i.e.,
bandages, clothing, etc., before adhering the patch. Placing the user patch
partially or wholly over a bandage or other obstruction could cause skin
irritation or tissue damage during stimulation.
5 Firmly adhere the user patch to the skin, making sure that the patch
is in full contact with the skin.
If the user patch is not firmly adhered
to the skin and moves, stimulation may become uncomfortable or
ineffective.
Turning On the Remote Control
Note:
The remote control must be within 10 feet of the external pulse
transmitter when the system is activated to ensure optimal radio
communication. If the components are not within 10 feet of each other, the
radio frequency icon will flash red.
Remote Control Off Remote Control On
The remote control on/off button is located at the top of the remote
control front panel.