User's Manual Part 5
109
Chapter 9 - Maintenance and Cleaning
Disinfecting the L300 Plus Components
All NESS L300 Plus System Kit components excluding the L300 FS Cuff and
the Thigh FS Cuff straps may be disinfected using CaviWipes™ (Metrex,
Orange, CA) per the manufacturer’s instructions. Metrex products are sold
through authorized dealers worldwide. If you have difficulty locating Metrex
CaviWipes, please call the Bioness Client Relations Department at (800)
211-9136, Option 3.
To disinfect the L300 Plus System components:
1. Clean the surface that faces the skin with a wet CaviWipes disinfection
wipe.
Note: Read the manufacturer's instructions for use, and follow standard
precautions for personal protection as appropriate.
2. Using one or more new CaviWipes, wipe the surface again, thoroughly,
for 3 minutes. The surface should be visibly wet.