User Manual Part 1

13
Chapter 3 - Environmental Conditions that Affect Use
Electromagnetic Emissions
The NESS L300 Plus System needs special precautions regarding
electromagnetic compatibility (EMC). The system needs to be installed and
put into service according to the EMC information provided in this manual.
See Appendix.
The NESS L300 Plus System was tested and certified to use the following:
x DC power supply as provided by Bioness Inc, manufactured by FRIWO,
Part No. FW7555M/05.
x
W cable (3-way splitter) as provided by Bioness Inc, Model No. L3P-5A00.
Manufactured by Tamuz Electronics Ltd.
Warnings
x
Do not use the NESS L300 Plus System within three feet of shortwave or
microwave therapy equipment. Such equipment may produce instability
in the output of the L300 and Thigh RF Stim Units.
x Remove the NESS L300 Plus System before undergoing any diagnostic
or therapeutic medical procedure such as x-ray examination, ultrasound,
Magnetic Resonance Imaging (MRI), etc.
x
The NESS L300 Plus System should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the equipment or
system should be observed to verify normal operation in the configuration
in which it will be used.
x
The use of accessories, transducers, and cables other than those specified,
with the exception of transducers and cables sold by the manufacturer of
the NESS L300 Plus System as replacement parts for internal components,
may result in increased emissions or decreased immunity of the NESS
L300 Plus System.