Chapter 10 - Electronically Registering New Components
Registration (Important: Read Steps 1–4 Before Starting)
1. Turn off the Control Unit.
2. Simultaneously press and hold for three seconds the mode and
minus buttons on the Control Unit. The Control Unit will beep when
registration begins.
3. The Control Unit digital display will show two alternating GREEN
arches while registration is in process.
4.
Within 15 seconds of initiating the registration procedure, repeatedly
press and release the pressure sensor.
5.
When registration is complete, (ā€œCā€ for complete) will appear in the digital
display, the Gait Sensor indicator will turn GREEN for a few seconds, and
the Control Unit will beep. If registration fails, wait 20 seconds for the RF
Stim Units to enter energy-saving mode, and then repeat the procedure.
Figure 74: Registering a new Gait Sensor.
Press and hold the
mode and minus
buttons for three
seconds.
New
Unregistered
Gait Sensor