Chapter 10 - Electronically Registering New Components
10
Electronically Registering New Components
When a NESS L300 Plus Control Unit, an RF Stim Unit, or a Gait Sensor is
replaced, the new component must be electronically registered to the other
NESS L300 Plus components for the system to communicate wirelessly.
NOTE: Components can only be successfully registered once. Additional
attempts will elicit an error indication.
Registering a New L300 Plus Control Unit
Setup
1.
Connect the new L300 Plus Control Unit and RF Stim Units to the system
charger set during registration.
2. Place the FS Cuffs with RF Stim Units attached, the Gait Sensor, and the
new L300 Plus Control Unit close together on a table but not touching.
3.
Turn off the old Control Unit and put it in an envelope for shipping to Bioness.
Then, place it at least 30 feet from the NESS L300 Plus components you
are registering.
4. Make sure all other NESS L300 Plus System components are at least 30
feet from the NESS L300 Plus components you are registering.
Registration
1. Turn off the new Control Unit.