Chapter 8 - Operating the L300 Plus Control Unit
Turning on Audio Feedback During Stimulation
The NESS L300 Plus System can be configured to provide audio feedback
during stimulation. You can turn on the audio feedback feature for each RF
Stim Unit individually using the RF selection buttons and the mute button.
Turning on Audio Feedback for the L300 FS Cuff
To turn on audio feedback during stimulation for the L300 FS Cuff:
1. Place the Control Unit in standby mode.
2. Press and hold the "down" button (to select the L300 RF Stim Unit)
and the mute button for three seconds. Both the Control Unit and the
L300 RF Stim Unit will emit an audio alert when stimulation turns on.
To turn off the audio feedback, press the mute button or the on/
off button.
Turning on Audio Feedback for the Thigh FS Cuff
To turn on audio feedback during stimulation for the Thigh FS Cuff:
1. Place the Control Unit in standby mode.
2. Press and hold the "up" button (to select the Thigh RF Stim Unit)
and the mute button for three seconds. Both the Control Unit and the
Thigh RF Stim Unit will emit an audio alert when stimulation turns on.
To turn off the audio feedback, press the mute button or the on/
off button.