71
Chapter 8 - Operating the L300 Plus Control Unit
Operating the L300 Plus Control Unit
RF Communication Safety Features
The L300 Plus Control Unit, L300 RF Stim Unit, Thigh RF Stim Unit, and
Gait Sensor must be within RF communication range of each other and their
batteries charged for the L300 Plus System to operate. If the components
become separated, or if a battery is discharged, RF communication will be lost.
If a communication failure occurs that involves the L300 RF Stim Unit:
• The RF Communication Indicator on the Control Unit will flash RED.
• The L300 RF Stim Unit indicator on the Control Unit will be solid RED.
• The status light on the L300 RF Stim Unit will flash RED.
• The Control Unit and L300 RF Stim Unit will emit an audio alert.
• The NESS L300 RF Stim Unit will deliver a warning default stimulation
to lift the foot for six seconds before the L300 Plus System shuts down.
If a communication failure occurs that involves the Thigh RF Stim Unit:
• The RF Communication Indicator on the Control Unit will flash RED.
• The Thigh RF Stim Unit indicator on the Control Unit will be solid RED.
• The status light on the Thigh RF Stim Unit will flash RED.
• The Control Unit and Thigh RF Stim Unit will emit an audio alert.
• The remaining NESS L300 Plus System components will continue to
operate.