User Manual

37
Chapter 5 - System Components and Accessories
L300 Cloth Electrodes
The L300 cloth electrodes are used to deliver stimulation from the L300 RF
Stim Unit to the lower leg muscles. See Figure 11. They are designed to be
skin friendly and provide an alternative for people who do not prefer the L300
hydrogel electrodes. They are also for people who have sensitivities to the
L300 hydrogel electrodes or other known sensitivities (for example, allergy
or skin sensitivity to tape/adhesive).
The L300 FS Cuff uses two cloth electrodes. Each one snaps to a cloth
electrode base.
The L300 cloth electrodes must be wet before use and after every three to
four hours of use. If the L300 FS Cuff is removed for more than one hour, the
cloth electrodes need to be rewet. Allow the cloth electrodes to dry between
uses. Do not use the L300 FS Cuff without electrodes.
Cloth Electrode
Figure 11: L300 cloth electrode, front (left) and back (right).
Caution: Change the electrodes every two weeks.
Caution: Only use electrodes provided by Bioness Inc.