Problem Solution
The "bad electrode
contact" indication
appears.
1. Turn OFF the system.
2. Remove the orthosis from your leg.
3. Make sure the electrode protective
covers were removed and there is
no disturbance to contact between
the electrodes and the skin.
4. Make sure that the electrodes
are installed properly according
to section "Replacing
Electrodes"), and that a
snapping sound was heard when
the electrode was inserted into
its base.
5. The electrodes may be missing or
damaged if this is the case then
install new electrodes according to
section "Replacing Electrodes"
6. Put the Orthosis on your calf,
turn the Control Unit ON
again, and press the Trigger
button.
If the problem persists then:
7. Check that the orthosis electronic
module is in place and locked it (see
the section Removing and Installing
the Orthosis Electronic Module), if
not reattach it to the orthosis.
If the problem persists then contact your
NESS L300 distributor.