89
Chapter 11 - Maintenance and Cleaning
System Kit Carrying Case
The NESS H200 Wireless System Kit carrying case may be cleaned and
low-level disinfected using CaviCide
(Metrex, Orange, CA) or 70% isopropyl
alcohol (IPA) per the following instructions:
CaviCide:
1. Spray the entire surface of the System Kit carrying case with CaviCide.
2. Use a clean towel to remove any surface contaminants. Soil, etc., will
impede the disinfectant's effectiveness, if not removed.
3.
Spray the entire surface of the System Kit carrying case again with CaviCide.
4. Keep spraying the entire carrying case surface as needed to keep it wet
for 10 minutes.
70% IPA:
1. Wipe the entire surface of the System Kit carrying case with a cloth or
wipe saturated with 70% IPA.
2. Use a new cloth or wipe saturated with 70% IPA to remove any surface
contaminants. Soil, etc., will impede the disinfectant's effectiveness, if
not removed.
3. Wipe the entire surface of the System Kit carrying case again with a new
cloth or wipe saturated with 70% IPA.
4. Use additional new cloths or wipes saturated with 70% IPA as needed to
keep the entire surface of the carrying case wet for 10 minutes.