The Orthosis (except for the wrist insert) may be cleaned and low-level
disinfected using wipes or cloths saturated with 70% isopropyl alcohol (IPA)
per the instructions below:
1.
Use one saturated disinfectant wipe or cloth to thoroughly wet the component
surface.
2. Use a second saturated disinfectant wipe or cloth to remove any surface
contaminants. Soil, etc., will impede the disinfectant's effectiveness, if
not removed.
3. As needed, use additional saturated disinfectant wipes or cloths to keep
the components surface wet for 3 minutes.
Note: Be sure to follow the Bioness instructions for the speciļ¬ed contact time
to ensure an effective bacteria kill.
Wrist Insert
The wrist insert cannot be disinfected. The wrist insert can only be cleaned
with soap and water. Do not use 70% IPA on the wrist insert. If infection is a
concern contact your clinician or the Bioness Client Relations Department
at (800) 211-9136, Option 3, for assistance.