User's Manual
70
User's Guide
Testing Stimulation in the Orthosis
The stimulation test button is for testing whether the Orthosis is positioned
correctly on the arm. The button is located on the side of the Control Unit.
See Figure 9-1.
To test stimulation in the Orthosis:
1.
Make sure the system is in standby mode. The trigger button should
not be lit.
2. Press and hold the
stimulation test button to test stimulation of the
extensor muscles. (Stimulation will turn on and stay on until the button is
released.) When stimulation is on, the
trigger button will FLASH
YELLOW RAPIDLY.
3. Release the
stimulation test button to turn off stimulation.
4. Repeat to test stimulation of the flexor muscles.
5. If the Orthosis is not positioned correctly:
• Turn off the system.
• Open the Orthosis wing and remove the Orthosis. Do not slide
the Orthosis on your arm.
• Recheck the cloth electrodes to make sure they are positioned
correctly and wet.
• Put the Orthosis back on.
• Close the Orthosis extensor wing until no clicking can be
heard.
• Retest stimulation.