Chapter 3 - Environmental Conditions that Affect Use
•
Do not apply stimulation in the presence of electronic monitoring equipment
(e.g., cardiac monitors, ECG alarms), which may not operate properly
when the electrical stimulation device is in use.
•
Remove the NESS H200 Wireless System before undergoing any diagnostic
or therapeutic medical procedure such as x-ray examination, ultrasound,
Magnetic Resonance Imaging (MRI), etc.
•
The use of accessories, transducers, and cables other than those specified,
with the exception of transducers and cables sold by the manufacturer
of the NESS H200 Wireless System as replacement parts for internal
components, may result in increased emissions or decreased immunity
of the NESS H200 Wireless System.
•
The use of the accessory, transducer, or cable with equipment and systems
other than those specified may result in increased emissions or decreased
immunity of the NESS H200 Wireless System.
•
The NESS H200 Wireless System may be interfered with by other equipment,
even if that other equipment complies with CISPR (International Special
Committee on Radio Interference, International Electrotechnical Commission)
emission requirements.
• Do not use the NESS H200 Wireless System within three feet of short
wave or microwave therapy equipment. Such equipment may produce
instability in the stimulation output by the H200 Wireless Orthosis.