User's Manual
119
Chapter 15 - Appendix - EMI Tables
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note: All calculations were made according to tables 204 and 206 of IEC 60601-1-2
for equipment that is not life-supporting using factors of 3.5 in 0.15–800 MHz and 7 in
800–2500 MHz. There are no requirements for ISM bands in these tables.