User's Manual
116
User's Guide
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The NESS H200 Wireless System is intended for use in the electromagnetic environment
specified below. The customer or the user of the NESS H200 Wireless System should
assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment—
Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the NESS H200 Wireless
System, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Conducted
RF
IEC 61000-
4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to
80 MHz
Recommended separation
distance:
d = 1.2√P
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz to
2.5 GHz
[E
1
] = 3 V/m
in 80 MHz to
2.5 GHz
[E
1
] = 10 V/m
in 26 MHz to
1 GHz
Recommended separation
distance:
d = 0.4√P, 80–800 MHz range
d = 0.7√P, 800–1000 MHz range
d = 2.3√P, 1000–2500 MHz range