User's Manual

113
Chapter 15 - Appendix - EMI Tables
15
Chapter
Appendix - EMI Tables
Guidance and Manufacturers Declaration Electromagnetic Emissions
The NESS H200 Wireless System is intended for use in the electromagnetic environment
specified below. The customer or the user of the NESS H200 Wireless System should
assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment —
Guidance
RF emissions
CISPR 11
Group 1 The NESS H200 Wireless System
uses RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to cause
any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The NESS H200 Wireless System is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies