H200 Wireless Orthosis Pulse Parameters
Pulse Balanced Biphasic
Waveform Symmetric
Intensity (Peak) 0–80 mA, 1-mA resolution (positive phase)
Maximum Current Intensity
(rms)
Electrodes #1, #2, #3, #5: 13.1 mA rms
Electrode #4: 18.6 mA rms
Max Voltage 120 V
Symmetric
Positive Pulse Duration (μsec) 100 200 300
Negative Pulse Duration (μsec) 100 200 300
Inter-Phase Interval (μsec) 50
Max Total Pulse Duration
(μsec)
250 450 650
Load Range 0–5000 ohm (Subject to max voltage limitation)
Nominal Load 500 ohm
Max Power Load 500 ohm (80 mA, 120 V)
Pulse Repetition Rate 20–45 Hz, 5-Hz resolution
Ramp Up 0–3.1 seconds
Ramp Down 0–3.1 seconds
Max. Duration of Stimulation
Program
4 hours, 5-minute resolution