The Control Unit does not light up when turned on.
The Control Unit needs to be charged.
If the problem persists, contact the Bioness Client Relations
Department, Option 3.
I received a replacement Control Unit. When I turn it on, the RF
communication indicator FLASHES RED and a "U" appears in the digital
display. The Orthosis status light and stimulation light are not lit.
A replacement component needs to be electronically registered to the
existing component for the NESS H200 Wireless System to communicate
wirelessly.
For instructions on how to register a replacement component,
see "Electronic Registration of Replacement Parts."
I tried the electronic registration procedure, but I never saw the alternating
arches in the digital display. The replacement component is not working.
Turn off the Control Unit.
Press the minus button and the trigger button on the Control Unit,
and then press the trigger button on the Orthosis, to start the registration
process.