Chapter 13 - Troubleshooting
When I turn on the H200 Wireless Control Unit, it beeps and the RF
communication indicator FLASHES RED
. The status light and
stimulation light on the Orthosis are not lit.
RF communication has failed. The Orthosis battery most likely is discharged.
Connect the system charger. (The NESS H200 Wireless
System can be charged during operation.) When
communication is restored, the RF communication indicator
will stop flashing and the
status light on the Orthosis will
FLASH GREEN.
I hear a beep and the status light on the Orthosis FLASHES RED.
A faulty electrode contact error has occurred. One or more of the cloth
electrodes is not in contact with the skin.
Turn off the Control Unit and remove the Orthosis.
Thoroughly cleanse the skin, removing any oils.
Remove and examine the cloth electrodes. If the cloth
electrodes are old or damaged, then replace them.
Clean the electrode bases, if necessary.
Wet and replace the cloth electrodes.
Make sure the extensor wing is closed properly.