Chapter 13 - Troubleshooting
13
Chapter
Troubleshooting
If you have any questions or concerns, please call the Bioness Client Relations
Department at (800) 211-9136, Option 3.
RF Communication Failure
The Orthosis and Control Unit communicate wirelessly. If RF communication
fails, the Control Unit RF Indicator will FLASH RED and emit an audio alert.
RF Failure Alert Problems/Solutions
Control Unit
FLASHES RED
Radio Communication Failure
• Make sure the H200 Wireless Orthosis and the Control
Unit are within 10 feet of each other.
• If the components are within range and working properly,
then turn the Control Unit off and back on.
Reorient the Control Unit.
• Check for obstructions or sources of interference.
• Change the H200 Wireless cloth electrodes.
• Contact the Bioness Client Relations Department.