User's Manual

Page
12
of
14
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Guidance and manufacturers declaration electromagnetic emission
The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The 2006-2B Blood Pressure
Monitor uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B
The 2006-2B Blood Pressure
Monitor is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC/EN61000-3-2
Not applicable
Voltage fluctuations
/flicker emissions
IEC/EN61000-3-3
Not applicable
Guidance and manufacture’s declaration–electromagnetic immunity
The 2006-2B Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure
that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV
contact
15 kV air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
±2kV for power
supply lines
±1kV for
input/output lines
Not
applicable
Not applicable
Surge
IEC/EN61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not
applicable
Not applicable
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC/EN61000-4-11
<5%UT(>95% dip
in UT)
for 0.5cycle
40%UT(60%dip
in UT)
for 5 cycles
70%UT(30%dip
in UT)
for 25 cycles
<5%UT(>95% dip
in UT)
for 5s
Not
applicable
Not applicable