Use Instructions

Version: BBN-LBL-IFU-FULL Rev. 0.5 8
Technical Specifications
Product Name BioButton
Model Number BIOST03020
Battery 140mAh, CR1632
Heart Rate
*
Range 40 to 125 beats per minute (<± 5 beats per minute)
Respiratory Rate
*
Range 10 to 30 breaths per minute
Skin Temperature Range 86°F to 107.6°F (30°C - 42°C)
Skin Temperature Accuracy < 96.4°F ± 0.5°F (< 35.8°C ± 0.3°C)
96.4°F to 98°F ± 0.3°F (35.8°C to 37°C ± 0.2°C)
98°F to 102°F ± 0.2°F (37°C to 39°C ± 0.1 °C)
102°F to 106°F ± 0.3°F (39°C to 41°C ± 0.2°C)
> 106°F ± 0.5°F (> 41°C ± 0.3°C)
Ambient Temperature 50°F to 110°F ± 1°F (10°C to 43°C)
STORAGE CONDITIONS
Temperature Range -4°F to 122°F (-20°C to 50°C)
Humidity Range 0 - 95% RH
OPERATING CONDITIONS
Temperature Range 32°F to 122°F (0°C to 50°C)
Humidity Range 0 - 95% RH
Atmospheric Pressure 70 - 102 kPa
Water Resistance IP47
COMMUNICATIONS
Communication Technology Bluetooth (BT4.2)
Distance Max. 10 meters (30 feet) line of sight
Radio Modulation GFSK Guassian frequency shift keying
Radio Frequency 2.4 – 2.5 GHz
Transmit Power 0dBm
Security AES-CTR 128 bit encryption (Advanced encryption standard counter mode)
*
Measurements are taken at rest