User's Manual
10 11
2.1 Inserting/exchanging batteries
The Biocomfort blood glucose meter is de
-
livered with a 3V lithium cell [type CR2032],
which must be inserted before taking the first
measurement.
If the battery symbol appears on the display,
this is an indication that the battery is already
weak and has to be exchanged soon. If the
battery symbol is blinking on the display, this
indicates that the battery must be exchanged.
A weak battery is to be exchanged so that er
-
ror-free measuring results can be guaranteed.
When the battery is exchanged, all settings
which have been made on the device are
retained, as well as the stored measurements.
Insert battery
Important information regarding the
control solution:
Please observe the separate user instructions
for the control solution.
– Always shake the control solution well
before use.
– The control solution can be damaged if
exposed to heat. For this reason, you
should immediately seal the vial immedi-
ately after use and keep it in a cool place.
– The control solution may not be used
past the expiration date. Once the vial
is opened, it can be used for 3 months.
When you open it for the first time,
write down the date on the vial.
Getting started
2
2 How to start up your
gluco-comfort blood glucose
meter
+
CR2032
gluco-comfort
Modell:BGM10
5
REF:590020SN:0526002A
IVDzurEigenanwendung
zurausschließlichenVerwendungmit
TeststreifenREFBGS105
r.
F.:20%-85%Batt.:CR-20323V
BiocomfortDiagnosticsGmbH&Co.KG
BernhäuserStraße17
D73765Neuhausena.d.Fildern
biocomfort
0197
10
ºC
40
ºC
+
CR2032
Additional important information concern-
ing the test strips:
Please observe the separate user instructions
for the test strips.
– The test strips could be damaged by effects
of temperature and light. For this reason,
they should be used immediately after
being removed from the test strip con-
tainer.
– The test strips may not be used past the
expiration date. The test strips can be used
four months after the container has been
opened. When you open the container for
the first time, write down the date on the
container.
– Close the container carefully after remov-
ing the test strip. Always leave the contain-
er closed and keep it in a cool, dry place.