User's Manual

48 49
Note: If you are not able to solve an occurring problem,
contact Biocomfort Diagnostics GmbH & Co. KG.
7
Error
Measures Remarks
23 error
(BGM 105)
The wireless connection
could not be cut.
Switch off the device and
restart.
Measurement cannot be carried
out.
24 error
(BGM 105)
Communication error Only occurs with the software,
which is available as an acces
-
sory.
32 error Function test of the system
was not completed success
-
fully.
Switch off the device and
restart.
If this error message appears
repeatedly, have the device
checked by the manufacturer.
error and
blinking bat
-
tery symbol
Battery is almost completely
used up.
Exchange the battery and
restart the device.
When the battery symbol blinks,
the blood glucose meter cannot
be used.
message
Diabetes experts have suggested that glucose
meters should agree within 15mg/dL of a
Causes
laboratory method when the glucose concen-
tration is lower than 75 mg/dL, and within
chart below displays how often the gluco-
comfort meter achieves this goal. The chart
is based on a study done on 100 patients to
20% of a laboratory method when the glucose
concentration is 75 mg/dL or higher. The
laboratory results.
For glucose results lower than 75 mg/dL, the
see how well the gluco-comfort compared to
match the laboratory method within 15 mg/dL
is 96% (39/40).
percentage (and number) of meter results that
that match the laboratory method within 20%
For glucose results at 75 mg/dL or higher,
is 97% (60/60).
Note: When meter results are compared to
the percentage (and number) of meter results
the laboratory results, results below 75 mg/dL
are compared in mg/dL.
7 Performance data summary
Performance data summary