User's Manual

40
41
Attention:
If the levels are not reached, please repeat
your measurements. If the values are still out
-
side the specified tolerances, please contact
the manufacturer of the blood glucose meas
-
uring system. Do not use the blood glucose
measuring system until it has been made sure
that the displayed values are correct.
Note:
Attention:
If the measuring cell of the test strip has
direct contact with the tip of the control
solution bottle, the enzyme of the measuring
cell could get into the control solution,
rendering it useless.
To keep the control measurement from being
included in the average calculation of the
blood glucose values, you must mark the
control measurement as such.
To do this, press the MEM button when the
meter is ready for measurement (Step 1 in
Section 3.4). In the display, the abbreviation
ctl” (control) appears to the right of the drop
symbol. The changeover to control mode is
confirmed with a beep. The function of the
individual measuring ranges can be checked
with the different control solutions:
Normal blood glucose levels are simulated
with the gluco-comfort control solution
GCS105.
Low blood glucose levels are simulated
with the gluco-comfort control solution
GCS205.
High blood glucose levels are simulated
with the gluco-comfort control solution
GCS305.
Notes:
Please follow the separate user instructions
for the control solutions.
Shake the control solution before use.
The control solution may not be used past
the expiration date.
Store the control solution at room tempera
-
ture.
The control range printed on the test
strip packaging only applies for the
control solution. It is not a recommended
range for your blood glucose value!
The control solution is not a cleaning
solution for the meter.
5.4 Evaluating the control measurement
When the measuring system is working
correctly, the measured blood glucose
values must correspond with the ranges
which are listed on the test strip packaging
for normal, low and high.
gluco-comfort฀฀
Kontroll-Lösung฀GCS฀101
12ml
Normal฀฀฀zur฀Eigenanwend฀฀g
•฀vor฀der฀Verwendung฀gut฀s฀฀ütteln
•฀nicht฀zum฀Verzehr฀geeign฀฀t
•฀nicht฀verbrauchte฀Kontroll฀ngist30Tage
฀฀฀nach฀Anbruch฀der฀Ampul฀฀฀verwerfen:
•฀Anbruchdatum:
Biocomfort฀Diagnostics฀Gm฀&
Co.฀KG
Bernhäuser฀Straße฀17
D฀73765฀Neuhausen฀a.฀d.฀Fil฀
gluco-comfort
mem
biocomfort


mem
Function check and care
5
Use only gluco-comfort control solutions
(GCS105, GCS205 or GCS305), since only
these are adapted to the measuring principle
of the Biocomfort blood glucose meter and
gluco-comfort test strip combination.
monitored by a healthcare provider.
We recommend that you periodically
compare this test system to another test
system known to be well maintained and