SPECIFICATIONS
Model no.:TD-4202
Dimension&Weight: 96mm(L)x45mm(W)x23mm(H), 76.15g
Power source: two 1.5V AAA alkaline batteries
Display: LCD
Memory: 450 measurement results with respective dates and times
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time count-down
Auto shutdown after 3 minutes of idleness
Temperature warning
Operating condition: 10°C~40°C, below 85% R.H. (noncondensing)
Meter Storage/Transportation condition: -20°C to 60°C, below
95%R.H.
Measurement units: Either mg/dL or mmol/L
Measurement range: 20~600mg/dL (1.1~33.3mmol/L)
The specications may be changed without prior notice.
Performance Characteristics
Accuracy: ±15mg/dL when glucose <75mg/dL
±20% when glucose 75mg/dL
Precision: This study shows the CV (correlation variation)
is less than 5%.
The device has tested to meet the following standards:
98/79/EC, IEC 60601-1, IEC 61010-1, IEC 60601-1-2, IEC61326,
and ISO 15197