Owner manual

| 11SsoAdvanced
Universal SYBR
®
Green Supermix Instruction Manual | 11
Dynamic Range
Determine the general trend of the slope where linearity (R
2
) and efciency are within acceptable
ranges, as specified above.
Sensitivity
Determine the lowest concentration of the serial dilution where replicate reproducibility is high
and the R
2
of the standard curve is ≥0.980.
Specificity
Evaluate the melt profiles when using SYBR
®
Greenbased assays. A single sharp peak with a
T
m
close to the calculated T
m
should be present.
Fig. 8. Example of a highly specific SYBR
®
Green assay where the melt curve exhibits a single
melt peak around 8587°C indicating a single PCR amplicon present.
1,200
60 70 80
d(RFU)/dt
90
1,000
800
600
400
200
0