User guide
Introduction
4  |  S3 Biosafety System
The front panel of the S3 biosafety system has two access doors (Figure 4). 
■n
 Sample access door — provides access to the loading stage touch screen lock, and  
sort chamber 
■n
 Fluidics access door — provides access to bulk fluidics door and allows fluidics  
containers to be exchanged
All four vinyl side panels are connected to the aluminum frame using magnetic attachment. Both 
access doors are attached to the front panel by magnets. The panels and doors can be detached 
using minimal effort and are designed for easy accessibility to the S3 or S3e cell sorter inside. 
CAUTION: Risk of danger!
Magnets could affect the functioning of pacemakers and implanted heart defibrillators. 
■n
A pacemaker could switch into test mode and cause illness
■n
A heart defibrillator may stop working
While the magnets used on this biosafety system are not high powered, care should 
be taken for those who may be sensitive. 
■n
If you wear these devices keep sufficient distance from magnets until tested to be safe
■n
 Warn others who wear these devices that they should stay a safe distance from  
the magnets 
Specifications
Table 5. S3 biosafety system specifications.
 Power requirements  100–240 VAC, 50–60 Hz, 1.25 Amp
 Power and efficiency  125 W
 Heat load  50 W as <85 BTU/hr 
 Airflow  175 CFM (cubic feet per minute)
 Blower  R1G 175 DC motorized impeller
 Fuse  2.0 A slow blow
 Weight  52 lb 
 Dimensions (W x H x D)  135 x 146 x 96 cm, 53 x 57.5 x 37.75 in
 Sound level  65 dB at maximum airflow 
 Operating temperature  18–25°C
 Operating humidity  20–60%
 HEPA filtration  99.99% or better on 0.3 µm- IEST-RP-CC-034 
 HEPA Filters  23.625 x 23.625 x 3.54 in, certify annually
 Top panel  6.4 mm polycarbonate 
 Base floor  19 mm high density polyethylene 
 Vinyl sides  30 mil PVC film and 18 oz vinyl coated polyester 
 Frame  25 mm anodized aluminum T-slot profile 
Fig. 4. Front panel access doors.
Sample door
Fluidics door