QX100 Droplet Generator Instruction Manual2 |
Chapter 1 QX100 Droplet Generator
1.2 QX100 Droplet Generator
The QX100 droplet generator uses microfluidics to combine oil and water (sample) to create the droplets 
required for ddPCR analysis. It processes up to eight samples at a time, in under 2 minutes. 
Following sample preparation using either the ddPCR supermix or One-Step RT-ddPCR kit for probes,  
20 μl each of eight prepared samples (or blanks) and droplet generator oil are transferred to the droplet 
generator (DG
™
) cartridge. The loaded cartridge is covered with a gasket and placed in the QX100 droplet 
generator. There, the samples and oil are combined within the microchannels of the cartridge to create an 
emulsion of ~20,000 monodisperse, nanoliter-sized droplets for each of the samples. Following droplet generation, 
the droplets are transferred to a standard 96-well PCR plate and amplified to end point using a standard thermal 
cycler. 
When cycling is complete, the plate is loaded into the QX100 droplet reader. The droplet reader sips each 
sample, singulates the droplets, and streams them in single file past a two-color detector. The detector reads 
each droplet and determines which contain a target (+) and which do not (–).
The QX100 droplet generator includes the components listed in Table 1.1. Additional requirements for droplet 
generation and PCR are listed in Table 1.2. For complete system requirements, refer to the QX100 Droplet 
Reader Instruction Manual (bulletin 10026321).
QX100 droplet generator.
Table 1.1. QX100 droplet generator components. Catalog # refers to replacement items (quantities may be different).
Component  Description  Catalog # 
QX100 droplet generator  Instrument used for droplet generation  186-3002
DG8 droplet generator  Microfluidic cartridge used to mix sample and oil to  186-3008
cartridges and gaskets (24)  generate droplets; gaskets seal the cartridge to 
  prevent evaporation and apply pressure required  
  for droplet formation 
Droplet generator cartridge  Position and hold the droplet generator cartridge    186-3051 
holders (2)  in the instrument for droplet generation
Power cord  Connects QX100 droplet generator to power source  Call technical support