Accessories User Manual
63
7.3.2 Error Codes
Error
code
Error Cause Recovery
0000
No error
8071
Invalid Command
The received data is not a command for the
instrument
8072
Parameter out of range The received parameter is out of range
8073
Device Not In Remote
Mode
The instrument is not in remote mode
8074
Device busy The instrument is doing a job.
8075
Filter wheel Filter wheel jammed. “reset” command.
8076
8077
Light bulb burned out Insufficient light level “reset” command.
8078
8079
8080
8081
8082
8083
Incubator error Incubator over heat.
8084
Plate carrier error-1 Plate carrier home sensor closed status error “reset” command.
8085
Plate carrier error-2 Plate carrier home sensor open status error “reset” command.
8086
Plate carrier error-3 Plate carrier jammed before reading “reset” command.
8087
Plate carrier Position Plate carrier jammed after reading “reset” command.
8088
Light level error
There is too big difference of light between
channels
“reset” command.
8089
Filter missing Light is exceeding allowable range “reset” command.
8090
Disagreement of security
status
The security status from P.C was disagreement.
8091
RS232C error RS232C hardware error
8092
A/D converter error Malfunction of A/D converter. “reset” command.
8093
Measurement stop by stop
key
The user input “stop” key during measurement. “reset” command.