Accessories Manual
16
Error Message Probable Cause Recommended Solution
Xe flash
The Xenon Lamp did not flash. If this occurs only a few times a year,
it is no problem.
Hi-Voltage power supply
damaged.
Contact Bio-Rad.
Xenon Flash Lamp damaged. Contact Bio-Rad.
Gain Mes-1
Measuring electronics
damaged.
Contact Bio-Rad.
Analog offset voltage is too
high.
Contact Bio-Rad.
Gain Mes-2
Selected filter is not installed. Install filter.
Measuring electronics
damaged.
Contact Bio-Rad.
Gain Mes-3
Xenon Flash Lamp damaged. Contact Bio-Rad.
Measuring electronics
damaged.
Contact Bio-Rad.
Optical Block is out of the
starting point.
Contact Bio-Rad.
Lenses and Mirrors may be
dirty.
Contact Bio-Rad.
Gain Ref-1
Measuring electronics
damaged.
Contact Bio-Rad.
Analog offset voltage is too
high.
Contact Bio-Rad.
Gain Ref-
Measuring electronics
damaged.
Contact Bio-Rad.
Gain Ref-3
Xenon Flash Lamp damaged. Contact Bio-Rad.
Measuring electronics
damaged.
Contact Bio-Rad.
Lenses and Mirrors may be
dirty.
Contact Bio-Rad.
A/D Mes.
Malfunction of A/D converter. Contact Bio-Rad.
A/D Incu
Malfunction of A/D converter. Contact Bio-Rad.
EEPROM
EEPROM chip trouble. Contact Bio-Rad.