Software Updates User guide
Plates
50
5. Control sample tab. Click on one sample to use as a control for relative gene 
expression calculations.
6. Click OK to apply the information entered to your plate. Any additional edits can be 
made using the Plate Editor. 
NOTE: Returning to the Auto layout tab will reset the plate layout. The layout can 
also be reset by selecting To o l s  > Clear Plate.
TIP: To read text within wells of the plate layout more easily, click the Zoom plate 
box for a magnified view.
Select Fluorophores Window
The Select Fluorophores window lists fluorophores that can be selected to load into the Plate 
Editor well loading controls. To open the Select Fluorophores window, click the Select 
Fluorophores button on the right side of the Plate Editor. 
NOTE: The fluorophores listed depend on the scan mode; when SYBR/FAM only is 
chosen, only channel 1 fluorophores are shown in the Select Fluorophores window.
NOTE: You cannot add or remove fluorophores in this list; you must calibrate the 
new fluorophores on an instrument in the Calibration Wizard (page 151). After 
calibration, the new fluorophore is added to the Select Fluorophore window.
Click the Selected checkbox next to the fluorophore name to add or remove fluorophores to 
the list on the right side of the Plate Editor window.
In this example, several dyes are selected from the list of available fluorophores (Figure 33).
Figure 33. Select Fluorophores window.
• Click the Color box next to the fluorophore name and select a new color to represent 
each fluorophore in the Plate Editor window and Data Analysis charts
NOTE: Before beginning the run, the software verifies that the fluorophores you 
specified in the plate are calibrated on that instrument. You cannot run a plate if it 
includes fluorophores that have not been calibrated on that instrument.










