Software Updates User guide

CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch Systems Instruction Manual
141
NRT (no reverse transcriptase control) with a C
q
less than xx. Input a C
q
cutoff value
Positive control with a C
q
greater than xx. Input a Cq cutoff value
Unknown without a C
q
Standard without a C
q
Efficiency greater than xx. Input a reaction efficiency cutoff value that is calculated
for the standard curve
Efficiency less than xx. Input a reaction efficiency cutoff value that is calculated for
the standard curve
Std Curve R^2 less than xx. Input a cutoff R^2 value for the standard curve
Replicate group C
q
Std Dev greater than xx. Input a cutoff standard deviation that is
calculated for each replicate group
Custom Export Tab
Select the Custom Export tab (Figure 98) to define the default settings for the fields that will be
exported and their export format when the Custom Export option is chosen.
Figure 98. Custom Export tab in the User Preferences window.
File export formats include text (*.txt), CSV (*.csv), Excel 2007 (*.xlsx), Excel 2003 (*.xls), XML
(*.xml), and HTML (*.html)
The following items can be chosen for export:
Sample Description. Well, Fluorophore, Target Name, Content, Replicate #, Sample
Name, Biological Set Name, and Well Note
Quantification. C
q
, Starting Quantity, C
q
Standard Deviation, and Quantity Standard
Deviation
Melt Curve. Melt Temperature, Peak Height, Melt Peak Begin Temperature, and End
Temperature
End Point. End Point Call and End RFU
Customize Column Names. Rename column headers