Software Updates User guide
CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch Systems Instruction Manual
97
The End Point tab shows the average RFU values to determine whether or not the target was 
amplified by the last (end) cycle. Use these data to determine if a specific target sequence is 
present (positive) in a sample. Positive targets have higher RFU values than the cutoff level you 
define. 
TIP: To create an end point protocol, open the Protocol tab (Run Setup window) 
and select Options > End Point Only Run.
The software displays these data in the End Point tab:
• Settings. Adjust data analysis settings
• Results. Shows the results immediately after you adjust the Settings
• Well Selector. Select the wells with the end point data you want to show
• Well spreadsheet. Shows a spreadsheet of the end RFU collected in the selected wells
Figure 73. Layout of the End Point analysis tab.
The Results list includes this information:
•Lowest RFU value. Lowest RFU value in the data
• Highest RFU value. Highest RFU value in the data
• Negative Control Average. Average RFU for the wells that contain negative controls
• Cut Off Value. Calculated by adding the tolerance (RFU or Percentage of Range listed in 
the Settings) and the average of the negative controls. Samples with RFUs that are 
greater than the cutoff value will be called “Positive.” To adjust the cutoff value, change 
the RFU or Percentage of Range
The Cut Off Value is calculated using this formula:
Select a tolerance by one of these methods:
• RFUs (default). Select this method to use an absolute RFU value for the tolerance. The 
minimum RFU tolerance value is 2. The maximum is the absolute value of the highest 
RFU value minus the absolute value of the lowest RFU value. The default RFU tolerance 
value is 10% of the total RFU range
• Percent of Range. Select this method to use a percentage of the RFU range for the 
tolerance. The minimum percent of range is 1%. The maximum percent of range is 99%. 
The default percent of range is 10%
Cut Off Value Negative Control Average  Tolerance+=