Software Updates Instruction Manual

MiniOpticon Instruction Manual
79
Select Percentage of Range to view the data as a percentage of the RFU range
Select wells in the well selector to focus on subsets of the data
Select a well group (page 57) to view and analyze a subset of the wells in the plate.
Select each well group by name in the Well Group pull-down menu in the toolbar
Data Description for End Point Analysis
Table 31 lists the types of information shown in the spreadsheet in the End Point tab.
Allelic Discrimination Tab
The Allelic Discrimination tab assigns the genotypes to wells with unknown samples using the
RFU or C
q
of positive control samples (Figure 58). Use this data to identify samples with
different genotypes, including Allele 1, Allele 2, Heterozygote, Unknown, Control 1, or Control
2.
NOTE: The data for allelic discrimination must come from multiplex runs with at
least two fluorophores. Each fluorophore identifies one allele in all samples.
Allelic discrimination analysis requires the following minimal well contents:
Two fluorophores in each well, except the wells that contain positive controls can
contain only one fluorophore
One fluorophore that is common to all wells in the well group
NTC (no template control) samples if you want to normalize the data
The software displays allelic discrimination data in these layouts:
RFU or C
q
chart. View the data in a graph of RFU or Cq for Allele 1/Allele 2. Each point
in the graph represents data from a single fluorophore in one well
Well spreadsheet. Shows a spreadsheet listing the allelic discrimination data collected
in each well of the plate
Well selector. Select the wells with the end point data you want to show
Table 31. End Point spreadsheet contents
Information Description
Well Well position in the plate.
Fluor Fluorophore detected.
Content A combination of the Sample type and Replicate #.
End RFU RFU at the end point cycle.
Call Positive or Negative, where positive samples have an RFU
value greater than the average RFU of the negative controls
plus the Cut Off Value.
Sample Sample Name loaded in the Plate Editor.