Software Manual
3   |   System Control
78    |    NGC Chromatography Systems and ChromLab Software
6. (Optional) Select the Enabled checkbox for SIM 2 and repeat steps 5a—5d for a 
second external detector.
7. Click OK.
To disable the connection
 On the Device Input tab, clear the Enabled checkbox for the device and click 
OK.
Note:  Disabling the SIM connection does not turn off the detector.
Device Output Tab
You can convert digital NGC data to analog voltage via the SIM and output the 
voltage to an external recording device, such as a light scattering detector. The SIM 
converts and outputs NGC trace information as voltage through its digital-to-analog 
converter (DAC) channels. When enabled, the external device can receive the 
converted trace information. For the device you can define its
 Signal type
 Minimum and maximum output range (from 0–10 volts)
 Minimum and maximum unit range 
To enable output to external devices
1. Ensure that the SIM is connected to the NGC instrument.
2. On the back of the SIM, ensure that the device is connected to DAC A.
3. Select File > System Settings to open the System Settings dialog box.
4. Choose the Device Output tab.