Software Manual
7   |   Evaluating Results
254    |    NGC Chromatography Systems and ChromLab Software
4. In the section, enter the values for bed height and particle diameter to calculate 
N/L, HETP, or rHETP.
Tip:   For predefined columns, you can change the prepopulated values for 
bed height and particle diameter in this section without changing the 
values in the Method Editor
.
5. Click Apply to calculate the selected column performance statistics and display 
them in the table.
Tip:   You can include the column performance statistics in the Run report. 
See Run Reports on page  277 for information about creating Run reports.
Resetting the Column Performance Properties
To reset the properties to their default values
 Click Reset to reset the properties checkboxes to their default values.
Deleting Column Performance Data
To delete column performance data for the displayed trace
 Click Delete.
Note:  This also removes the Column Performance table.