Software Manual
6   |   Creating a Method
190    |    NGC Chromatography Systems and ChromLab Software
Special Considerations
Rules for generating scout sequences can vary depending on the parameter chosen 
to scout. This section provides a list of considerations to take into account when 
generating a scout sequence.
Generating %B Scout Sequences
 You can scout on two or more steps with the same starting %B values. If 
you select two or more steps with different starting values, you must clear 
the steps with the different values.
 When a linear gradient is selected for scouting and the value for the next or 
previous %B step is the same as the selected gradient, that step is 
automatically selected for scouting. Clear that step if it is not required.
Note:  If an isocratic step is selected for scouting, no other steps are 
automatically selected. Y
ou can choose others manually. 
 When the starting value for %B is 100, the increments (on Page 3) default 
to a negative value.
Generating pH Scout Sequences
 This option is available only when the fluidic scheme includes the buffer 
blending valve and buffer blending is selected in Method settings.
 The minimum and maximum pH scouting range is defined by the buffer 
selection made in Method settings.
Generating Duration Scout Sequences
 Choosing two or more steps that have different starting values or that have 
different units is not supported. Clear the steps with different starting 
values or units to proceed.