Creating a Method from Standard Phases
User Guide   |   169
Serpentine is selected by default. Fractions can also be collected by rows or 
columns for microplates. Possible choices appear for the rack you select.
If you have outlet valves in the fluidic scheme, choose a starting valve and port 
and modify the fraction size.
8. Under Flow, specify the default flow rate for the method. You can change the 
default flow rate within individual phases if needed.
Select the Control the flow to avoid overpressure checkbox if you want the flow 
rate to decr
ease and the method to continue running in the event that the 
system r
eaches high pressure limits. If you do not select this checkbox the 
system pumps will stop, pausing the method.
If you have an air sensor installed and you want to monitor and stop the system 
pumps if air is detected in the lines, select the Detect end of buf
fer with Air 
Sensor checkbox.
Note:  If the method pauses due to an end-of-buf
fer signal, the fluidic lines 
must be primed befor
e r
esuming the method. Because the system pump’s 
Purge button is inactive in Method mode, you must reprime the pumps 
manually. See Priming and Purging the System on page  94 for information 