Software Manual
6   |   Creating a Method
162    |    NGC Chromatography Systems and ChromLab Software
Multicolumn Tandem Multicolumn tandem purification uses these templates 
when two different chromatography techniques are 
combined into one method. First, samples are injected, 
bound to a column and washed to remove any main 
contaminants. Next, the target fractions are eluted and 
either applied directly onto another column in tandem or 
stored temporarily in a sample loop or container. If stored, 
they are then re-injected onto a second column. In either 
case, the second column is then eluted and the purified 
fractions are collected with the BioFrac fraction collector 
or an outlet valve. 
Reverse Phase  Reverse phase chromatography utilizes hydrophobic regions 
on target proteins to bind to a hydrophobic column matrix. A 
gradient of increasing organic solvent concentration/ratio is 
used to elute the proteins from the column matrix. 
Size Exclusion (Gel Filtration) Size exclusion/gel filtration chromatography separates 
pr
oteins based on their molecular size or weight. Proteins do 
not bind to the column matrix and are typically eluted 
isocratically. Select a buffer system that maximizes the 
stability of the target protein.
System Test These methods test the performance of NGC systems that 
have a UV detector
. They are intended to assess the flow rate 
and gradient accuracy of the system pumps, the functionality 
of the mixer, and the responses of the UV and conductivity 
detectors. In most cases, a visual examination of the resulting 
chromatogram can confirm proper performance. 
Table 6.  Standard Method Templates, continued
Method Template Explanation










