Software Manual
Standard Method Templates
User Guide   |   161
Chromatofocusing  Chromatofocusing chromatography uses a charged matrix to 
bind protein molecules. A pH gradient is used to elute the 
bound proteins, which elute when the pH is the same as the 
pI of the protein of interest (overall charge = 0). Special buffer 
systems are required to perform the pH gradient over a large 
range.
Desalting Desalting is usually used for buffer exchange. Proteins do not 
bind to the column matrix and ar
e typically eluted isocratically 
in the void volume of the column. Select a buffer system that 
maximizes the stability of the target protein.
Hydrophobic Interaction  Hydrophobic interaction chromatography uses high salt 
buf
fers to adsorb target proteins to a hydrophobic column 
matrix. Decreasing salt concentrations are then used to elute 
and separate bound proteins. 
Mixed Mode Mixed mode chromatography uses a column matrix with 
hydr
ophobic and charged ionic interactions. Proteins can be 
eluted using a gradient of pH (eluting when the pH = pI of the 
target protein) or salt (increasing salt to elute from the 
charged moiety or decreasing salt to elute from the 
hydrophobic moiety of the column matrix).
Multicolumn Sequential Multicolumn sequential purification uses these templates 
when 
multiple samples must be purified on multiple 
columns. The samples are injected sequentially either by 
using a sample pump with sample inlet valve or through 
sample loops. Each sample is loaded onto a column and 
washed to remove contaminants that can cause sample 
degradation. The columns are then eluted using either 
step or linear gradient protocols in a sequence. The 
fractions are collected with the BioFrac™ fraction 
collector or an outlet valve. 
Table 6.  Standard Method Templates, continued
Method Template Explanation