Software Manual
6   |   Creating a Method
160    |    NGC Chromatography Systems and ChromLab Software
Standard Method Templates
The workflow for standard method templates consists of the following phases:
 Equilibration
 Sample application
 Column wash
 Elution
 Column wash
 Re-equilibration for the next run
ChromLab includes the following eight standard method templates. 
Table 6.  Standard Method Templates 
Method Template Explanation
Affinity Affinity chromatography is the separation of biomolecules 
based on highly specific interactions. For example: 
antibody/antigen or antibody/protein A, chelation 
(polyhistidine-tag/nickel), enzyme/substrate 
(glutathione-s-transferase/glutathione).
Anion Exchange  Anion exchange chromatography uses a positively charged 
column matrix to bind negatively char
ged protein molecules. 
Proteins are eluted from the matrix using a gradient of 
increasing ionic strength (typically NaCl). Select a buffer 
system with a pH lower than the pI of the target protein to 
enhance protein binding.
Cation Exchange  Cation exchange chromatography uses a negatively charged 
column matrix to bind positively char
ged protein molecules. 
Proteins are eluted from the matrix using a gradient of 
increasing ionic strength (typically NaCl). Pick a buffer system 
with a pH higher than the pI of your target protein to enhance 
protein binding.