Phase Controls and Parameters
User Guide   |   139
Loading the Loop with a Sample Pump
When a sample pump is used to load the loop, you can define the flow rate, the 
volume to load into the loop, and, subsequently, the volume of sample to be 
injected onto the column. You also have the option to interrupt loading the loop if air 
is detected. If you select this checkbox and air is detected in any of the lines 
connected to air sensors set to detect end of sample, the system stops the sample 
pump and the method continues to the next step.
Fluidic Scheme Includes a Sample Inlet Valve
If your fluidic scheme includes at least one sample inlet valve and you choose to 
load the loop through the sample pump, you have the option to perform a 
pr
e-injection wash with buffer to flush the sample pump and lines and the loop. If 
you select this checkbox, you can choose the port on the sample inlet valve from 
which to draw buffer solution. You can define the flow rate and the volume of buffer 
for the preinjection wash.
You also have the option to flush the sample pump and related connections after 
sample injection. Again, you can choose the port on the sample inlet valve fr
om 
which to draw buf
fer solution and define the flow rate and the volume of buffer for 
the postinjection wash.