Software Manual
Method Settings Parameters
User Guide   |   133
BioFrac Settings — by default, fraction collection begins in the first tube of the 
first rack. The fraction size is the default for the rack type. 
BioFrac Collection Pattern — by default, Serpentine is selected. For 
micr
oplates, fractions can also be collected by rows or columns.
Outlet Valve Settings — by default, fraction collection begins fr
om 
Outlet valve 1 (O1) Port 2. O1 Port 1 connects to the BioFrac™ fraction 
collector if it is pr
esent or directs flow to waste if the fraction collector is absent. 
The default fraction size is 50 ml.
Flow
Flow Rate — shows the default rate of flow for the method.
Control the flow to avoid overpressure — monitors the pr
essure and reduces 
the flow rate by 50% when the pressure gets within a certain percentage of the 
maximum (default is 80%, see Control Flow Tab in Chapter 3 for more 
information).
Detect end of buffer with air sensor — stops the system pumps if air is 
detected in any of the lines connected to air sensors set to detect end of buf
fer. 
Air sensor settings are determined on the Air Sensors tab in the System 
Settings dialog box. (See Air Sensors Tab in Chapter 3 for more information).
Buffer Selection
Note:  When the configuration does not include a buffer blending valve, buffer 
settings are inactive. 
You can choose one of three Buffer Selection modes:
 Manually Prepared — buffers are user prepared and fed directly into the 
pumps or selected by means of buffer inlet valves.
 Manually Prepared via Blending Valve — buffers are user prepared and 
fed to the pump through buffer blending valve ports Q1 and Q4. Gradients 
are formed by the blending valve rather than the pumps. Because of this, 
both pumps can run at their full capacity simultaneously, which doubles 
the available flow rate range.










