4    |    Performing a Manual Run
106    |    NGC Chromatography Systems and ChromLab Software
6. Double-click the Sample Pump module, if installed, and choose the following 
settings in the dialog box that appears:
a. If one or two sample inlet valves are in line, select a port from which to 
draw sample.
b. Under Flow Control, select the flow rate.
c. Under Pressure Limits, select the lower and upper pressure limits.
d. (Optional) Select Control Flow to avoid overpressure. This setting causes 
the flow rate to decr
ease automatically if the system exceeds set pressure 
limits.
e. If not already set, click Change Position and select the position for the 
inject valve.
f. (Optional) Select Enable Air Sensor if air sensors are in line and set to 
detect end of sample.
Important:  Select flow control settings for the run before you click Start. 
Clicking Start starts the run.