Software Manual
4    |    Performing a Manual Run
98    |    NGC Chromatography Systems and ChromLab Software
3. To purge an individual line of air bubbles and fill it with buffer:
a. Select a port to open.
b. (Optional) Change the default flow rate. The default flow rate is set to 
20  ml/min.
c. Click Start.
d. Repeat steps 3a–c for any other port to purge.
4. To purge all lines of air bubbles and fill them with buffer:
a. Click All.
b. (Optional) Change the default flow rate. The default flow rate is set to 
20  ml/min.
c. Click Start.
5. Monitor the pump pressure and outlet flow to ensure that the fluid is flowing 
pr
operly.
The system pumps and the buffer blending valve start operating at the 
specified flow rate, exchanging the solution in the inlet lines and expelling any 
trapped bubbles, cycling sequentially thr
ough the ports Q1 to Q4. After the 
specified duration, the pumps stop momentarily while the blending valve 
switches to port Q3 in order to flush the lines with water before closing the 
valve. No change is allowed during this process for the specified duration.
Note:  You can stop the cycle before it completes by clicking Stop. Before 
stopping the pur
ge cycle, ensure that all buffer lines are filled with the 
appropriate buffers before starting a run.
Gradient via 
Blending Valve 
(High Flow)
Buffer blending valve ports Q1 and Q4 are used as inlets for 
buffers A and B. Gradients are formed by the buffer blending 
valve rather than by the pumps. This enables each pump to 
run at their full flow rate capacity which doubles the total 
flow rate, for example to 20 ml/min for F10 pumps.
Mode Description










