Chapter 1: CFX Automation System II Installation
| 13CFX Automation System II | 13
Calibrate the Automation System
The CFX Automation System II must be properly calibrated during installation. Recalibrate any
time the system is moved or CFX Systems are added or removed.
Default calibrations for your CFX System configuration are automatically loaded during the
system configuration step. The automation software includes a calibration wizard that aids in
making the fine adjustments required for proper calibration. To calibrate the system:
■n
Select Tools > Calibration
■n
Select the positions to calibrate
Fig. 12. Platform configuration window.
Fig. 13. Base serial number locations.
Serial number located on the
base of the instrument.
Home screen on CFX System
touch screens. Base serial number
is located on this screen.